PolarisQ Installation Requirements

Supplies

  • Power (minimum): Four dedicated outlets each rated at 120 VAC +6%/-10%, 50/60 Hz, 15 A or 230 VAC +/- 10%, 50/60 Hz, 16 A.      
  • Helium: purity 99.999% with less than one ppm each of water, oxygen, and total hydrocarbons      
  • CI reagent gases: methane, isobutane or ammonia with purity 99.99%

Environment

Complete system averages 3940 W (13,450 Btu/h) output when considering air conditioning needs.  Operating environment must be 15-31°C (59-88°F) and relative humidity must be 40-80% with no condensation.  Optimum operating temperature is 18-26°C (65-78°F).

System Dimensions/Weights

Complete GC/MS system requires 2 m (6 ft) of workbench space. Allow 16 cm (6 in) of clearance behind the instrument.

PolarisQ Analyzer

  • MS: 44 cm height x 33 cm width x 68 cm depth      
  • Weight: 45 kg (98 lb)

TRACE Gas Chromatograph

  • GC: 44 cm height x 61 cm width x 65 depth      
  • Weight: 55 kg (120 lb)

Data Station

  • Computer CPU: 42 cm height x 16.5 cm width x 44.5 cm depth      
  • Monitor: 42 cm height x 40 cm width x 43 cm depth      
  • Keyboard: 2.5 cm height x 46 cm width x 18 cm depth      
  • Total weight: 32 kg (70 lb)